Clinical Comparison of Contour and Applanation Tonometry and Their Relationship to Pachymetry FREE

Christoph Kniestedt, MD; Shan Lin, MD; Joyce Choe, MD; Alan Bostrom, PhD; Michelle Nee, MD; Robert L. Stamper, MD
[+] Author Affiliations

Author Affiliations: Departments of Ophthalmology (Drs Kniestedt, Lin, Choe, Nee, and Stamper) and Epidemiology and Biostatistics (Dr Bostrom), University of California–San Francisco; and Departments of Ophthalmology, Cantonal Hospital Winterthur, Winterthur, Switzerland (Dr Kniestedt), and University Hospital Zurich, Zurich, Switzerland (Dr Kniestedt).


Arch Ophthalmol. 2005;123(11):1532-1537. doi:10.1001/archopht.123.11.1532.
Text Size: A A A
Published online

Objectives  To compare intraocular pressure readings of recently introduced dynamic contour tonometry (DCT) with pneumatonometry (PTG) and Goldmann applanation tonometry (GAT) and to correlate central corneal thickness (CCT) with these readings.

Design  Prospective, cross-sectional observation and instrument validation study. We included 258 independent eyes with normal anterior segment examinations results, irrespective of glaucoma diagnosis or glaucoma suspect. After pachymetry, DCT, PTG, and GAT were performed in a randomized order. Intraocular pressures as measured by DCT, PTG, and GAT were compared with each other and with CCT.

Results  Eyes with thinner CCTs tended to yield lower intraocular pressure measurements by GAT. A significant correlation (Pearson product moment correlation, P<.001) between CCT and GAT was found with a regression of 0.25 mm Hg per 10 μm (R2 = 0.060). Variation of CCT had no significant effect on intraocular pressure measurements by PTG (P = .10; R2 = 0.01) and DCT (P = .80; R2<0.01). A piecewise regression model showed that GAT readings are not linearly correlated with CCT. Comparison of the slopes below and above 535 μm showed the highest significance (P<.001).

Conclusions  Goldmann applanation tonometry readings are potentially influenced by CCT, whereas PTG and DCT seem to be less dependent on CCT. Correlation between CCT and GAT is not linear. A simple correction formula suggesting a linear relationship might not be correct.

Figures in this Article

In 1955, Goldmann1 was the first to successfully apply the principle of corneal applanation for tonometry. Goldmann applanation tonometry (GAT) has since become the gold standard for noninvasive tonometry and is still the most popular method of measuring intraocular pressure (IOP). All types of applanation tonometers are governed by the Imbert-Fick law.2,3 This law states that when a thin, perfectly elastic sphere is flattened, the force per unit area applied to flatten the sphere and the force per unit area of the internal pressure should equalize each other. The flattened area and the subsequently displaced volume are supposed to be small in relation to the total area and volume of the sphere. The latter might be true in a human eye, but human sclera and cornea are neither perfectly elastic nor thin and flexible. Goldmann and Schmidt4 were well aware that corneal rigidity and, therefore, corneal thickness must oppose the effect of indentation and applanation. The inventors of GAT calculated that the surface tension drawing the tonometer tip onto the cornea would be counterbalanced by the resistance to applanation offered by the cornea at a diameter of exactly 3.06 mm. Goldmann and Schmidt47 performed all of their measurements on corneal thicknesses ranging from 500 to 520 μm, which they assumed to be normal and representative of almost all eyes with healthy corneas.

Recently, comprehensive multicenter studies have demonstrated that corneal thickness varies considerably within healthy subjects. Because of this, IOP may be incorrectly assessed and, thus, management of glaucoma may be adversely affected.

It would be helpful to have a tonometer that measures IOP directly (ie, determining the pressure rather than a force) and is not biased by individual characteristics of the cornea and the observer. Kanngiesser et al8 developed a method for transcorneal and continuous IOP measurement. They call it dynamic contour tonometry (DCT) because the pressure-sensitive tip is not planar but closely resembles the curvature of the cornea.

The detailed physical hypothesis and theoretical considerations about DCT are described elsewhere.8 Briefly summarized, the DCT tip has a radius that is slightly larger than that of an average human cornea. Kanngiesser et al8 determined empirically that the radius of curvature needed to be 10.5 mm to get accurate results and to fit on most corneas. A pressure sensor (diameter, 1.7 mm) is embedded in the shell-shaped tonometer tip (Figure 1). Forcing the central disc area of the cornea into the contour of the DCT tip allows the examiner to measure the pressure of the eye directly on the external surface of the cornea because, in the condition of matched contours, the pressure on both sides of the cornea is theoretically equal. The IOP recorded by DCT is defined as the mean diastolic IOP during the period when the tonometer was in contact with the eye.

Place holder to copy figure label and caption
Figure 1.

The examination is performed with the patient in a sitting position at the slitlamp. A, Dynamic contour tonometer tip inserted into a Goldmann applanation tonometer tip holder. B, Drum is set at 1 g (appositional force = 9.81 mN [milliNewton]).

Graphic Jump Location

Proper investigation with the novel DCT on human cadaver eyes showed better absolute and relative accuracy than GAT and pneumatonometry (PTG).9,10 The dependence of central corneal thickness (CCT) could not have been investigated in vitro. However, Kaufmann et al11,12 and Siganos et al13 reported that DCT seems to be less dependent on CCT than GAT or noncontact air-puff tonometry on normal eyes and on eyes after the laser-assisted in situ keratomileusis procedure, respectively.

This study was performed to collect the early clinical experience using DCT in a glaucoma-based single-center patient population and to compare its dependence on CCT with that of PTG and GAT.

The present prospective study included a random sample of consecutive patients with glaucoma and suspected glaucoma who consented to the study protocol. All participants were seen at the Department of Ophthalmology, University of California–San Francisco between November 1, 2002, and April 30, 2003, and gave written informed consent before enrollment. Eyes were excluded if they had any corneal disease or acquired irregularity. The study protocol was approved by the Committee on Human Research at the University of California–San Francisco (H10262-22264-01). We examined 509 eyes of 258 consecutive patients and 258 independent eyes were included in the study. To reduce variability, only the right eyes were chosen. Ten right eyes had to be excluded owing to corneal edema, penetrating keratoplasty, prosthesis, and phthisis bulbi. In these cases, the 10 left eyes met study criteria and were included in place of the right eyes.

Visual acuity measurement, pachymetry, GAT, and PTG were performed by a technician certified for the Ocular Hypertension Treatment Study who was masked to DCT readings. Goldmann applanation tonometry, PTG, and DCT were performed in a randomized order. One measurement of at least 10 heartbeats was taken for further analysis for DCT and PTG. To reduce variability, the mean of 2 readings was applicable for GAT analysis. The 2 GAT readings were acquired by the technician and by 1 of us (C.K., S.L., J.C., or R.L.S.). Goldmann applanation tonometry was calibrated weekly and performed in the manner originally described by Goldmann1 and Goldmann and Schmidt47 using a BQ 900 slitlamp (Haag Streit, Bern, Switzerland). If pulsating hemirings were noticeable, an average setting was chosen with horizontally symmetric oscillation to both sides. The model 30 classic pneumatonometer (Medtronic Inc, Minneapolis, Minn) was used for all PTG readings throughout the study. The standard deviation cutoff was set according to the manufacturer’s manual to get sufficiently reproducible readings. To avoid possible interobserver variability, which is assumed to be minimal but not yet determined for DCT, 1 observer was selected to perform DCT (C.K.). Dynamic contour tonometry and GAT were performed with the patient sitting in an upright position at the slitlamp (Figure 1). For DCT, the pressure-sensitive tip was inserted into a GAT tip holder in a manner similar to that for the GAT tip (Figure 1A). The GAT drum was set to 1 g following the inventors’ protocol (Figure 1B). Observation through the slitlamp microscope reveals a fluorescein ring rather than 2 hemicircles. The purpose of the fluorescein ring is to visualize and confirm that the DCT is appropriately centered on the corneal surface. The ring should be located in the midperiphery, evenly distributed in a concentric manner around the pressure sensor (Figure 2), indicating the area of contour matching.

Place holder to copy figure label and caption
Figure 2.

View through the slitlamp microscope while the examination is taking place. The miniaturized piezoresistive pressure sensor (1) (diameter, 1.7 mm) is built into the center of the concavity (radius, 10.5 mm) of the contact surface and is surrounded by an evenly distributed and concentrally located fluorescein ring (diameter, 3-7 mm) (2). The wire connection to the infrared control unit (3) is also shown.

Graphic Jump Location

The CCT was assessed as an average of 5 consecutive measurements using an ultrasound pachymeter (Humphrey Instruments, San Leandro, Calif). The speed of sound was adjusted at 1640 m/s according to the internationally accepted standard velocity for human corneas.

Statistical analysis was performed with a mixed-effects regression model using SAS software (SAS Institute Inc, Cary, NC). The model treated patients and their eyes as random effects and did not assume equal variability in the 3 devices. Associations between continuous and other ordered variables were examined using the Spearman nonparametric correlation (Spearman ρ). Nonparametric Kruskal-Wallis and Mann-Whitney tests were also used to examine associations between categorical variables and continuous or ordered outcomes. Analysis of variance was used to compare IOP readings in the 3 devices. A P value (Spearman, Kruskal-Wallis, and Mann-Whitney) of <.05 was defined as statistically significant.

The possibility of different linear relationships between IOP and CCT for different ranges of CCT was investigated using piecewise regression methods.14 The slope of the IOP on CCT is assumed to be b1 for CCT ≤ X0 and b2 for CCT > X0. We also assume that the 2 lines intersect at CCT = X0. Mathematically, this model can be written as follows:

IOP = a + [b1(X − X0)(1 − I[X])] + [b2I(X)(X − X0)],

where X = CCT, a = estimated CCT value when X = X0, and I(X) = 0 if X<X0 and I(X) =1 if XX0.

This model was fit using multiple regression, and the F statistic for testing H0 (b1 = b2) was computed. Large values of this statistic are evidence against the null hypothesis of equal slopes in the 2 CCT regions. To find the optimal cutoff, the value of X0 was systematically varied from 500 to 600 μm in steps of 1 μm.

A total of 258 eyes underwent evaluation. Sixty-six eyes were diagnosed as being glaucoma suspect, including 23 eyes with ocular hypertension. One hundred seventy eyes were diagnosed as having a form of open-angle glaucoma. This group included eyes with primary open-angle glaucoma (n = 125), normal-tension glaucoma (n = 23), congenital glaucoma (n = 1), juvenile glaucoma (n = 1), pseudoexfoliation glaucoma (n = 16), and pigmentary glaucoma (n = 4). Finally, 22 eyes were found to have angle-closure glaucoma. The population consisted of 95 male and 163 female patients with a mean age of 69 years (median, 71 years; range, 14 - 97 years). The ethnic distribution was 181 white, 39 Asian, and 16 African American patients, 18 patients of Hispanic descent, and 4 patients of Arab (n = 2) or native East Indian (n = 2) extraction. The mean ± SD CCT of the entire group was 545 ± 38 μm.

Intraocular pressure was recorded using GAT, PTG, and DCT in a randomized order. Mean ± SD IOP as measured by GAT was 16.0 ± 3.0 mm Hg (range, 3-27 mm Hg); by PTG, 17.1 ± 4.1 mm Hg (range, 5.0-28.5 mm Hg); and by DCT, 18.3 ± 4.2 mm Hg (range, 5.0-31.1 mm Hg).

There was no significant intradevice IOP difference detected among the 6 measurement orders (ADP [n = 41], APD [n = 54], DAP [n = 36], DPA [n = 40], PAD [n = 53], and PDA [n = 34], where A indicates GAT; D, DCT; and P, PTG). The Kruskal-Wallis P value was .21 for DCT, .27 for GAT, and .59 for PTG. A strong correlation between all 3 devices was found (r = 0.86 for DCT vs GAT; r = 0.87 for DCT vs PTG; and r = 0.87 for GAT vs PTG; P<.001 for any device comparison).

With analysis of variance, the overall test of equality of IOP in the 3 devices was very strongly rejected (F = 147.12; P<.001). Tukey tests of pairwise differences showed all 3 devices to be significantly different (at P<.05) from each other.

Intraocular pressure measured with GAT was significantly correlated with CCT (y = 0.025x + 2.70; R2 = 0.06; P<.001) with a 0.25–mm Hg change per 10-μm variation in CCT based on linear regression analysis (Figure 3A). With a P value of .10, PTG in contrast did not reach enough significance to be correlated with CCT (y = 0.011x + 11.07; R2 = 0.01) (Figure 3B). Intraocular pressure measured with DCT showed no significant correlation to CCT (P = .80; y = 0.002x + 17.34, R2<0.01) (Figure 3C). Linear regression analysis of each of the diagnosis subgroups showed similar results with comparable significance levels for each tonometric device.

Place holder to copy figure label and caption
Figure 3.

Central corneal thickness (CCT) in correlation to intraocular pressure (IOP) readings obtained by using Goldmann applanation tonometry (GAT) (A), pneumatonometry (PTG) (B), and dynamic contour tonometry (DCT) (C). The GAT shows the steepest slope (0.025), indicating a statistically significant correlation with CCT. Pneumatonometry shows less correlation, and DCT shows no correlation at all.

Graphic Jump Location

The possibility of different linear relationships between IOP and CCT for different ranges of CCT was investigated using a piecewise regression model.14 The value of the CCT cutoff that maximized the F statistic was found to be 535 μm for all 3 IOP measures (GAT, F1,255 = 6.24; DCT, F1,255 = 3.15; PTG, F1,255 = 3.39; significance level, 5.02) (Figure 4). Goldmann applanation tonometry showed a difference in the comparison of the linear regressions below 535 μm (slope, 0.047; P = .001) and above 535 μm (slope, −0.040; P = .06), which was significant (P<.001). However, the slope −0.040 is not significant owing to higher measurement errors above 535 μm. Both DCT (P = .19) and PTG (P = .08) showed no significant change in the slopes at any cutoff point. Models across the entire CCT range with higher-order terms showed no significant nonlinear effects.

Place holder to copy figure label and caption
Figure 4.

Testing of equal slopes (F ratio) and value of central corneal thickness (CCT). The greatest difference between the slopes was found at 535 μm (P<.001) for Goldmann applanation tonometry (GAT) (F = 6.24). Dynamic contour tonometry (DCT) and pneumatonometry (PTG) also showed a similar cutoff at 535 μm (F = 3.15 and F = 3.39, respectively). However, unlike GAT, the differences did not reach statistical significance. (Significance level [F ratio testing equal slopes], 5.02).

Graphic Jump Location

Central corneal thickness has become an important biometric factor and is an essential part of the evaluation of glaucoma. The quality of pachymeters has changed considerably during the past few decades. At present, ultrasonic pachymeters have replaced the older optical pachymeters, which have been shown to be less accurate and measure consistently lower than ultrasonic pachymeters.1517 In their meta-analysis, Doughty and Zaman18 found a chronological upward trend in the reported averages for CCT during a 30-year period that is thought to be due to the change from optical to ultrasonic measuring methods. The group-averaged value for CCT using optical pachymetry was 525 μm (median), and for ultrasonic pachymetry, 544 μm (median). Thus, Goldmann and Schmidt’s47 value for CCT of 500 to 520 μm, which is based on optical means, might be, in fact, approximately 520 to 540 μm.

Therefore, recently published data are based on ultrasound pachymetry. These data showed mean CCTs of 552 μm,19 518 μm,20 554 μm,21 536 μm (a primary open-angle glaucoma sample), and 592 μm (an ocular hypertension sample).22 The Ocular Hypertension Treatment Study reported a mean CCT of 573 μm,23 and the Rotterdam Study described 537 μm24 with a very wide range of 193 μm. Central corneal thickness appears to be thicker in patients with ocular hypertension, which may be explained, in part at least, by the fact that some of these eyes are misclassified owing to IOP overestimation.2123 Argus25 described a mean ± SD CCT in his ocular hypertension group of 610 ± 33 μm. Subjects with primary open-angle glaucoma show a slightly thinner CCT than that of control subjects.21,22,24,25 Finally, normal-tension glaucoma is reported to be associated with CCT in the low 500-μm range.21,26

Numerous studies have been conducted to determine a correlation factor to define real IOP in eyes with unusually thin or thick corneas. Argus25 introduced a correction formula assuming 578 μm as normal. Stodtmeister27 published correction nomograms for applanation tonometry performed on corneas of a thickness different from 580 μm. Unfortunately, only a few studies are based on manometric measurements. Even if IOP was checked manometrically, caution in analyzing the results is necessary; for instance, Ehlers et al28 have compared CCT to a correction factor, derived from manometric IOP and applanating IOP, whereas Wolfs et al24 have plotted CCT to manometric values directly. Orssengo and Pye29 recently proposed a new nonlinear correlation formula. To our knowledge, its accuracy has not yet been proved with an independent manometric study. Furthermore, the fact that normal corneal thickness is assumed to vary widely from 450 to 600 μm makes it unclear at what CCT level to start using any nomogram.

Our study results emphasize the general suggestion that IOP as measured by GAT is dependent on CCT. Using a linear regression model, we found a significant correlation between GAT and CCT with a 0.25–mm Hg change per 10-μm variation in CCT. The average ± SD CCT of our glaucoma sample (545 ± 38 μm) and the correlation between CCT and IOP are clearly within the range indicated in most other studies (Table). Dynamic contour tonometry and PTG are not significantly correlated with CCT, although PTG is closer to a significant correlation (IOP variation of 0.11 mm Hg for every 10 μm). This may be clinically negligible in the CCT range obtained in this study. Dynamic contour tonometry showed the least correlation with CCT, with readings that were subject to change only 0.02 mm Hg for each 10 μm. However, the chosen study design has its limitations and its possible bias. Our population sample is based on patients with glaucoma, many of whom have far advanced disease, and we did not include a control group. For the comparison of the 3 devices, this might be irrelevant. The range of IOPs found in this population is somewhat limited because all of the patients with glaucoma were receiving pressure-lowering treatment.

Table Graphic Jump LocationTable. Relationship Between CCT and GAT

A review of the literature shows variations from 0.11 mm Hg21 to 0.71 mm Hg28,32,34 for every 10 μm of CCT change. These studies applied different study designs to different race and diagnosis groups; therefore, it is not surprising that they showed different mean CCTs. The fact that patient samples with different CCTs result in a wide range of correlation factors leads to the possibility that a linear correlation between IOP and the entire range of possible CCTs might not exist. We addressed this assumption with a piecewise regression model and have found that with thin corneas (<535 μm) the slope between CCT and IOP is significantly steeper than with normal or thick corneas. This analysis confirms our clinical experience using DCT that GAT’s underestimation with thin corneas is of a much greater issue than its overestimation with thick corneas. Taking only the slopes below and above the cutoff points at 20-μm steps, the correlation was always stronger and significant between thin CCTs (500, 520, and 540 μm) and IOP for GAT. Surprisingly, the slopes above the cutoff points turned out to be negative. However, this phenomenon was not significant owing to larger measurement errors on thick corneas. The reason for the higher measurement errors on thick corneas is not yet clarified. It is possible that thick corneas result in higher errors per se because corneal rigidity is increased, or that thick corneas may represent a nonhomogeneous group, some of which may be inherently thick while some may be thickened by subclinical edema. The latter would correspond to the negative correlation between relatively thick CCT and GAT, which was already observed by Simon et al.36

Dynamic contour tonometry is not significantly influenced by CCT and, therefore, the application of correction factors for unusually thin or thick corneas is unnecessary. Also, PTG appears not to be affected by CCT, whereas GAT is significantly influenced by CCT within the range investigated in this study. Goldmann applanation tonometry did not show a linear relationship to CCT.

Dynamic contour tonometry is a promising technology that may provide more accurate IOP measurements and, thus, allow better management of ocular hypertension and glaucoma. Further work is warranted to determine whether DCT keeps its reliability on abnormally thin corneas (eg, after a laser-assisted in situ keratomileusis procedure), differently hydrated corneas (eg, in the case of stromal edema), and corneas with irregular surfaces. Clinical studies that include manometric reference pressures would be necessary to address these questions appropriately.

Correspondence: Christoph Kniestedt, MD, Department of Ophthalmology, Cantonal Hospital Winterthur, Brauerstrasse 15, 8400 Winterthur, Switzerland (research@kniestedt.ch).

Submitted for Publication: April 29, 2004; final revision received December 2, 2004; accepted January 31, 2005.

Financial Disclosure: None.

Funding/Support: This study was supported in part by the Fund to Prevent and Fight against Blindness, Zurich, Switzerland, and That Man May See Foundation, San Francisco, Calif. Dr Kniestedt has received financial support from the Alfred Vogt Fund, Zurich.

Goldmann  H Un nouveau tonometre d’applanation. Bull Soc Ophtalmol Fr 1955;67474- 478
Imbert  A Theories ophthalmotonometres. Arch Ophthalmol 1885;5358- 363
Fick  A Ueber Messung des Druckes im Auge. Arch Physiol Hum Anat 1888;42633
Goldmann  HSchmidt  T On applanation tonography. Ophthalmologica 1965;15065- 75
PubMed Link to Article
Goldmann  HSchmidt  T Weiterer Beitrag zur Applanationstonmetrie. Ophthalmologica 1961;141441- 456
PubMed Link to Article
Goldmann  HSchmidt  T Studien mittels Applanationstonographie. Doc Ophthalmol 1966;20184- 213
PubMed
Goldmann  HSchmidt  T Ueber Applanationstonometrie. Ophthalmologica 1957;134221- 242
PubMed Link to Article
Kanngiesser  HEKniestedt  CRobert  YC Dynamic contour tonometry: presentation of a new tonometer. J Glaucoma 2005;14344- 350
PubMed Link to Article
Kniestedt  CNee  MStamper  RL Dynamic contour tonometry: a comparative study on human cadaver eyes. Arch Ophthalmol 2004;1221287- 1293
PubMed Link to Article
Kniestedt  CNee  MStamper  RL Accuracy of dynamic contour tonometry compared with applanation tonometry in human cadaver eyes of different hydration states. Graefes Arch Clin Exp Ophthalmol 2004;243359- 366
PubMed Link to Article
Kaufmann  CBachmann  LMThiel  MA Intraocular pressure measurements using dynamic contour tonometry after laser in situ keratomileusis. Invest Ophthalmol Vis Sci 2003;443790- 3794
PubMed Link to Article
Kaufmann  CBachmann  LMThiel  MA Comparison of dynamic contour tonometry with Goldmann applanation tonometry. Invest Ophthalmol Vis Sci 2004;453118- 3121
PubMed Link to Article
Siganos  DSPapastergiou  GIMoedas  C Assessment of the Pascal dynamic contour tonometer in monitoring intraocular pressure in unoperated eyes and eyes after LASIK. J Cataract Refract Surg 2004;30746- 751
PubMed Link to Article
Hawkins  D Point estimation of the parameters of piecewise regression models. Appl Stat 1976;2551- 57
Link to Article
Salz  JJAzen  SPBerstein  JCaroline  PVillasenor  RASchanzlin  DJ Evaluation and comparison of sources of variability in the measurement of corneal thickness with ultrasonic and optical pachymeters. Ophthalmic Surg 1983;14750- 754
PubMed
Giasson  CForthomme  D Comparison of central corneal thickness measurements between optical and ultrasound pachometry. Optom Vis Sci 1992;69236- 241
PubMed Link to Article
Gordon  ABoggess  EAMolinari  JF Variability of ultrasonic pachometry. Optom Vis Sci 1990;67162- 165
PubMed Link to Article
Doughty  MJZaman  ML Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach. Surv Ophthalmol 2000;44367- 408
PubMed Link to Article
Bhan  ABrowning  ACShah  SHamilton  RDave  DDua  HS Effect of corneal thickness on intraocular pressure measurements with the pneumotonometer, Goldmann applanation tonometer, and Tono-Pen. Invest Ophthalmol Vis Sci 2002;431389- 1392
PubMed
Gunvant  PBaskaran  MVijaya  L  et al.  Effect of corneal parameters on measurements using the pulsatile ocular blood flow tonograph and Goldmann applanation tonometer. Br J Ophthalmol 2004;88518- 522
PubMed Link to Article
Shah  SChatterjee  AMathai  M  et al.  Relationship between corneal thickness and measured intraocular pressure in a general ophthalmology clinic. Ophthalmology 1999;1062154- 2160
PubMed Link to Article
Bron  AMCreuzot-Garcher  CGoudeau-Boutillon  Sd’Athis  P Falsely elevated intraocular pressure due to increased central corneal thickness. Graefes Arch Clin Exp Ophthalmol 1999;237220- 224
PubMed Link to Article
Brandt  JDBeiser  JAKass  MAGordon  MO Central corneal thickness in the Ocular Hypertension Treatment Study (OHTS). Ophthalmology 2001;1081779- 1788
PubMed Link to Article
Wolfs  RCKlaver  CCVingerling  JRGrobbee  DEHofman  Ade Jong  PT Distribution of central corneal thickness and its association with intraocular pressure: the Rotterdam Study. Am J Ophthalmol 1997;123767- 772
PubMed
Argus  WA Ocular hypertension and central corneal thickness. Ophthalmology 1995;1021810- 1812
PubMed Link to Article
Morad  YSharon  EHefetz  LNemet  P Corneal thickness and curvature in normal-tension glaucoma. Am J Ophthalmol 1998;125164- 168
PubMed Link to Article
Stodtmeister  R Applanation tonometry and correction according to corneal thickness. Acta Ophthalmol Scand 1998;76319- 324
PubMed Link to Article
Ehlers  NBramsen  TSperling  S Applanation tonometry and central corneal thickness. Acta Ophthalmol (Copenh) 1975;5334- 43
PubMed Link to Article
Orssengo  GPye  D Determination of the true intraocular pressure and modulus of elasticity of the human cornea in vivo. Bull Math Biol 1999;61551- 572
Link to Article
Copt  RPThomas  RMermoud  A Corneal thickness in ocular hypertension, primary open-angle glaucoma, and normal tension glaucoma. Arch Ophthalmol 1999;11714- 16
PubMed Link to Article
Dohadwala  AAMunger  RDamji  KF Positive correlation between Tono-Pen intraocular pressure and central corneal thickness. Ophthalmology 1998;1051849- 1854
PubMed Link to Article
Ehlers  NHansen  FK Central corneal thickness in low-tension glaucoma. Acta Ophthalmol (Copenh) 1974;52740- 746
PubMed Link to Article
Whitacre  MMStein  R Sources of error with use of Goldmann-type tonometers. Surv Ophthalmol 1993;381- 30
PubMed Link to Article
Rosa  NCennamo  GBreve  MALa Rana  A Goldmann applanation tonometry after myopic photorefractive keratectomy. Acta Ophthalmol Scand 1998;76550- 554
PubMed Link to Article
Johnson  MKass  MAMoses  RAGrodzki  WJ Increased corneal thickness simulating elevated intraocular pressure. Arch Ophthalmol 1978;96664- 665
PubMed Link to Article
Simon  GSmall  RHRen  QParel  JM Effect of corneal hydration on Goldmann applanation tonometry and corneal topography. Refract Corneal Surg 1993;9110- 117
PubMed

Figures

Place holder to copy figure label and caption
Figure 1.

The examination is performed with the patient in a sitting position at the slitlamp. A, Dynamic contour tonometer tip inserted into a Goldmann applanation tonometer tip holder. B, Drum is set at 1 g (appositional force = 9.81 mN [milliNewton]).

Graphic Jump Location
Place holder to copy figure label and caption
Figure 2.

View through the slitlamp microscope while the examination is taking place. The miniaturized piezoresistive pressure sensor (1) (diameter, 1.7 mm) is built into the center of the concavity (radius, 10.5 mm) of the contact surface and is surrounded by an evenly distributed and concentrally located fluorescein ring (diameter, 3-7 mm) (2). The wire connection to the infrared control unit (3) is also shown.

Graphic Jump Location
Place holder to copy figure label and caption
Figure 3.

Central corneal thickness (CCT) in correlation to intraocular pressure (IOP) readings obtained by using Goldmann applanation tonometry (GAT) (A), pneumatonometry (PTG) (B), and dynamic contour tonometry (DCT) (C). The GAT shows the steepest slope (0.025), indicating a statistically significant correlation with CCT. Pneumatonometry shows less correlation, and DCT shows no correlation at all.

Graphic Jump Location
Place holder to copy figure label and caption
Figure 4.

Testing of equal slopes (F ratio) and value of central corneal thickness (CCT). The greatest difference between the slopes was found at 535 μm (P<.001) for Goldmann applanation tonometry (GAT) (F = 6.24). Dynamic contour tonometry (DCT) and pneumatonometry (PTG) also showed a similar cutoff at 535 μm (F = 3.15 and F = 3.39, respectively). However, unlike GAT, the differences did not reach statistical significance. (Significance level [F ratio testing equal slopes], 5.02).

Graphic Jump Location

Tables

Table Graphic Jump LocationTable. Relationship Between CCT and GAT

References

Goldmann  H Un nouveau tonometre d’applanation. Bull Soc Ophtalmol Fr 1955;67474- 478
Imbert  A Theories ophthalmotonometres. Arch Ophthalmol 1885;5358- 363
Fick  A Ueber Messung des Druckes im Auge. Arch Physiol Hum Anat 1888;42633
Goldmann  HSchmidt  T On applanation tonography. Ophthalmologica 1965;15065- 75
PubMed Link to Article
Goldmann  HSchmidt  T Weiterer Beitrag zur Applanationstonmetrie. Ophthalmologica 1961;141441- 456
PubMed Link to Article
Goldmann  HSchmidt  T Studien mittels Applanationstonographie. Doc Ophthalmol 1966;20184- 213
PubMed
Goldmann  HSchmidt  T Ueber Applanationstonometrie. Ophthalmologica 1957;134221- 242
PubMed Link to Article
Kanngiesser  HEKniestedt  CRobert  YC Dynamic contour tonometry: presentation of a new tonometer. J Glaucoma 2005;14344- 350
PubMed Link to Article
Kniestedt  CNee  MStamper  RL Dynamic contour tonometry: a comparative study on human cadaver eyes. Arch Ophthalmol 2004;1221287- 1293
PubMed Link to Article
Kniestedt  CNee  MStamper  RL Accuracy of dynamic contour tonometry compared with applanation tonometry in human cadaver eyes of different hydration states. Graefes Arch Clin Exp Ophthalmol 2004;243359- 366
PubMed Link to Article
Kaufmann  CBachmann  LMThiel  MA Intraocular pressure measurements using dynamic contour tonometry after laser in situ keratomileusis. Invest Ophthalmol Vis Sci 2003;443790- 3794
PubMed Link to Article
Kaufmann  CBachmann  LMThiel  MA Comparison of dynamic contour tonometry with Goldmann applanation tonometry. Invest Ophthalmol Vis Sci 2004;453118- 3121
PubMed Link to Article
Siganos  DSPapastergiou  GIMoedas  C Assessment of the Pascal dynamic contour tonometer in monitoring intraocular pressure in unoperated eyes and eyes after LASIK. J Cataract Refract Surg 2004;30746- 751
PubMed Link to Article
Hawkins  D Point estimation of the parameters of piecewise regression models. Appl Stat 1976;2551- 57
Link to Article
Salz  JJAzen  SPBerstein  JCaroline  PVillasenor  RASchanzlin  DJ Evaluation and comparison of sources of variability in the measurement of corneal thickness with ultrasonic and optical pachymeters. Ophthalmic Surg 1983;14750- 754
PubMed
Giasson  CForthomme  D Comparison of central corneal thickness measurements between optical and ultrasound pachometry. Optom Vis Sci 1992;69236- 241
PubMed Link to Article
Gordon  ABoggess  EAMolinari  JF Variability of ultrasonic pachometry. Optom Vis Sci 1990;67162- 165
PubMed Link to Article
Doughty  MJZaman  ML Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach. Surv Ophthalmol 2000;44367- 408
PubMed Link to Article
Bhan  ABrowning  ACShah  SHamilton  RDave  DDua  HS Effect of corneal thickness on intraocular pressure measurements with the pneumotonometer, Goldmann applanation tonometer, and Tono-Pen. Invest Ophthalmol Vis Sci 2002;431389- 1392
PubMed
Gunvant  PBaskaran  MVijaya  L  et al.  Effect of corneal parameters on measurements using the pulsatile ocular blood flow tonograph and Goldmann applanation tonometer. Br J Ophthalmol 2004;88518- 522
PubMed Link to Article
Shah  SChatterjee  AMathai  M  et al.  Relationship between corneal thickness and measured intraocular pressure in a general ophthalmology clinic. Ophthalmology 1999;1062154- 2160
PubMed Link to Article
Bron  AMCreuzot-Garcher  CGoudeau-Boutillon  Sd’Athis  P Falsely elevated intraocular pressure due to increased central corneal thickness. Graefes Arch Clin Exp Ophthalmol 1999;237220- 224
PubMed Link to Article
Brandt  JDBeiser  JAKass  MAGordon  MO Central corneal thickness in the Ocular Hypertension Treatment Study (OHTS). Ophthalmology 2001;1081779- 1788
PubMed Link to Article
Wolfs  RCKlaver  CCVingerling  JRGrobbee  DEHofman  Ade Jong  PT Distribution of central corneal thickness and its association with intraocular pressure: the Rotterdam Study. Am J Ophthalmol 1997;123767- 772
PubMed
Argus  WA Ocular hypertension and central corneal thickness. Ophthalmology 1995;1021810- 1812
PubMed Link to Article
Morad  YSharon  EHefetz  LNemet  P Corneal thickness and curvature in normal-tension glaucoma. Am J Ophthalmol 1998;125164- 168
PubMed Link to Article
Stodtmeister  R Applanation tonometry and correction according to corneal thickness. Acta Ophthalmol Scand 1998;76319- 324
PubMed Link to Article
Ehlers  NBramsen  TSperling  S Applanation tonometry and central corneal thickness. Acta Ophthalmol (Copenh) 1975;5334- 43
PubMed Link to Article
Orssengo  GPye  D Determination of the true intraocular pressure and modulus of elasticity of the human cornea in vivo. Bull Math Biol 1999;61551- 572
Link to Article
Copt  RPThomas  RMermoud  A Corneal thickness in ocular hypertension, primary open-angle glaucoma, and normal tension glaucoma. Arch Ophthalmol 1999;11714- 16
PubMed Link to Article
Dohadwala  AAMunger  RDamji  KF Positive correlation between Tono-Pen intraocular pressure and central corneal thickness. Ophthalmology 1998;1051849- 1854
PubMed Link to Article
Ehlers  NHansen  FK Central corneal thickness in low-tension glaucoma. Acta Ophthalmol (Copenh) 1974;52740- 746
PubMed Link to Article
Whitacre  MMStein  R Sources of error with use of Goldmann-type tonometers. Surv Ophthalmol 1993;381- 30
PubMed Link to Article
Rosa  NCennamo  GBreve  MALa Rana  A Goldmann applanation tonometry after myopic photorefractive keratectomy. Acta Ophthalmol Scand 1998;76550- 554
PubMed Link to Article
Johnson  MKass  MAMoses  RAGrodzki  WJ Increased corneal thickness simulating elevated intraocular pressure. Arch Ophthalmol 1978;96664- 665
PubMed Link to Article
Simon  GSmall  RHRen  QParel  JM Effect of corneal hydration on Goldmann applanation tonometry and corneal topography. Refract Corneal Surg 1993;9110- 117
PubMed

Correspondence

CME
Meets CME requirements for:
Browse CME for all U.S. States
Accreditation Information
The American Medical Association is accredited by the Accreditation Council for Continuing Medical Education to provide continuing medical education for physicians. The AMA designates this journal-based CME activity for a maximum of 1 AMA PRA Category 1 CreditTM per course. Physicians should claim only the credit commensurate with the extent of their participation in the activity. Physicians who complete the CME course and score at least 80% correct on the quiz are eligible for AMA PRA Category 1 CreditTM.
Note: You must get at least of the answers correct to pass this quiz.
You have not filled in all the answers to complete this quiz
The following questions were not answered:
Sorry, you have unsuccessfully completed this CME quiz with a score of
The following questions were not answered correctly:
Commitment to Change (optional):
Indicate what change(s) you will implement in your practice, if any, based on this CME course.
Your quiz results:
The filled radio buttons indicate your responses. The preferred responses are highlighted
For CME Course: A Proposed Model for Initial Assessment and Management of Acute Heart Failure Syndromes
Indicate what changes(s) you will implement in your practice, if any, based on this CME course.