Subbasal Nerve Density and Corneal Sensitivity After Laser In Situ Keratomileusis:  Femtosecond Laser vs Mechanical Microkeratome FREE

Sanjay V. Patel, MD; Jay W. McLaren, PhD; Katrina M. Kittleson, BS; William M. Bourne, MD
[+] Author Affiliations

Author Affiliations: Department of Ophthalmology, Mayo Clinic, Rochester, Minnesota.


Arch Ophthalmol. 2010;128(11):1413-1419. doi:10.1001/archophthalmol.2010.253.
Text Size: A A A
Published online

Objective  To compare changes in subbasal nerve density and corneal sensitivity after laser in situ keratomileusis (LASIK) with the flap created by a femtosecond laser (bladeless) vs a mechanical microkeratome.

Design  In a randomized paired-eye study, 21 patients received myopic LASIK with the flap created by a femtosecond laser in one eye and by a mechanical microkeratome in the fellow eye. Eyes were examined before and at 1, 3, 6, 12, and 36 months after LASIK. Central subbasal nerve density was measured by using confocal microscopy. Corneal mechanical sensitivity was measured by using a gas esthesiometer and was expressed as the ratio of mechanical threshold in eyes that received LASIK to mechanical threshold in concurrent control eyes.

Results  Subbasal nerve density and corneal sensitivity did not differ between methods of flap creation at any examination. Mean (SD) nerve density was decreased at 1 month (bladeless, 974 [2453] μm/mm2; microkeratome, 1308 [2881] μm/mm2) compared with the preoperative examination (bladeless, 10 883 [5083] μm/mm2, P < .001; microkeratome, 12 464 [6683] μm/mm2, P < .001) and remained decreased through 12 months (P < .001). Mechanical threshold ratios did not differ from that at the preoperative examination through 36 months for either LASIK treatment; when all LASIK eyes were combined, the mechanical threshold ratio was transiently higher (decreased sensitivity) at 1 month (1.29 [0.85]) compared with the preoperative examination (0.89 [0.73], P = .05).

Conclusions  The planar configuration of the femtosecond laser flaps is not associated with faster reinnervation compared with the microkeratome flaps. The prolonged decrease in subbasal nerve density after LASIK is not accompanied by a prolonged decrease in corneal sensitivity.

Trial Registration  clinicaltrials.gov Identifier NCT00350246

Figures in this Article

Corneal nerves are severed and ablated during laser in situ keratomileusis (LASIK), and regrowth of subbasal nerves is not complete several years after surgery.1 Denervation can result in a presumed neurotrophic epithelium with ocular surface symptoms and signs.2 LASIK flaps have traditionally been created with mechanical microkeratomes, but more recently, femtosecond laser technology has emerged as an alternative for flap creation3 and has been suggested to induce fewer signs of dry eye compared with microkeratomes.4 The very short duration of pulses delivered by femtosecond lasers enables precise corneal cuts of any geometric configuration. LASIK flaps created with femtosecond lasers are typically planar in contrast to flaps created by mechanical microkeratomes, which are thinner in the center than at the periphery and have tapered edges.5,6 The planar configuration of femtosecond laser flaps might allow improved realignment of the flap after LASIK compared with flaps created with mechanical microkeratomes and therefore confer an advantage for reinnervation and the return of corneal sensitivity.

In this randomized paired-eye study, we investigated whether the reinnervation of corneal subbasal nerves was faster after LASIK with a planar flap created by a femtosecond laser compared with LASIK with the flap created by a mechanical microkeratome. We also assessed changes in corneal sensitivity to mechanical stimuli during the first 3 years after surgery.

PARTICIPANTS

Twenty-one participants were recruited from patients attending the refractive surgery service at Mayo Clinic. All patients had myopia or myopic astigmatism, were 21 years or older, and were determined to be suitable candidates for LASIK after a rigorous screening examination. The mean (SD) participant age at surgery was 38 (10) years (range, 22-54 years). Twenty myopic, unoperated-on control participants who did not receive LASIK were also recruited and examined concurrently with those receiving LASIK. The mean (SD) age of the controls at their first examination was 38 (9) years (range, 21-51 years). Individuals were excluded if they had any corneal abnormalities; history of ocular disease, trauma, or surgery; or diabetes mellitus or other systemic disease known to affect the eye or if they used ocular medications. Systemic medications were permitted unless they were known to affect the cornea or anterior segment. This study complied with requirements of the Health Insurance Portability and Accountability Act and was approved by the Mayo Clinic institutional review board. Informed consent was obtained from all participants after explanation of the nature and possible consequences of the study.

RANDOMIZATION

Participants receiving LASIK were stratified by ocular dominance. One eye of each patient was randomized to receive LASIK with the flap created by a femtosecond laser and the other eye received LASIK with the flap created by a mechanical microkeratome.

LASIK PROCEDURE

Bladeless flap creation was performed with a 15-kHz femtosecond laser (IntraLase FS; IntraLase Corp, Irvine, California). All flaps were created to have a superior hinge with hinge angle of 45° or 55°, an intended thickness of 120 μm, and an intended diameter of 9.0 mm. Raster energy was 2.3 μJ, and side-cut energy was 2.5 μJ. The side-cut angle was 60° or 70° in all cases. Flaps created by the mechanical microkeratome (Hansatome; Bausch & Lomb, Rochester, New York) had a superior hinge with an intended thickness of 180 μm and an intended diameter of 8.5 or 9.5 mm. The stromal bed was ablated with an excimer laser (VISX Star S4; VISX, Santa Ana, California) at a radiant exposure of 160 mJ/cm2. Emmetropia was attempted in all cases by using an ablation zone that ranged from 6.5 × 6.5 mm for spherical corrections to 6.5 × 5.0 mm for astigmatic corrections. Postoperative topical medication regimens were identical for each eye and consisted of ciprofloxacin hydrochloride ophthalmic solution, 4 times daily for 5 days, and fluorometholone acetate, 0.1%, 4 to 8 times daily with a taper that extended across 3 weeks.

OUTCOME MEASURES

Patients were examined before LASIK and at 1, 3, 6, 12, and 36 months after surgery. Controls did not receive a surgical intervention and were examined concurrently with the LASIK patients but only at 0, 1, 12, and 36 months. At each examination, subbasal nerve density and mechanical corneal sensitivity were measured. Central flap thickness at 1 month was measured by using confocal microscopy as previously reported.7

SUBBASAL NERVE DENSITY

Subbasal nerve density was measured from confocal microscopy images acquired with a ConfoScan 3 confocal microscope (Nidek Technologies, Greensboro, North Carolina) for examinations before June 1, 2005, and with a ConfoScan 4 confocal microscope (Nidek Technologies, Freemont, California) for examinations from June 1, 2005, and afterward. Images were acquired by continuous through-focusing as described in detail previously.8 Briefly, after anesthetizing the cornea with topical proparacaine hydrochloride, 0.5%, an optical coupling medium (GenTeal Gel; Novartis Ophthalmics, East Hanover, New Jersey) was placed on the tip of the objective and z-ring adapter, and the operator advanced the objective and z-ring until the coupling medium contacted the cornea (the z-ring was available for the ConfoScan 4 confocal microscope only). The objective was aligned to obtain a centered image of the corneal endothelium; the focal plane was then advanced into the anterior chamber before initiating a through-focus examination of the full-thickness cornea from posterior to anterior. The focal plane was automatically reset to the starting position and the full-thickness scan was repeated. Video frames were recorded at 25 frames/s; the step distance between frames was 5 μm. Each cornea was scanned 2 to 4 times.

All images were reviewed, and the best image containing corneal subbasal nerves was manually selected. Randomized images of the nerves were given to a masked observer (K.M.K.), and the subbasal nerves were traced by using a semiautomated image analysis program (NeuronJ, a plug-in program to ImageJ9,10). The total length of corneal nerves within a central sample area of the image was measured, and nerve density was expressed as the length of corneal nerves per unit area. The sample area was 194 × 298 μm (horizontal × vertical) and was selected to minimize use of peripheral regions of the image where nerve visibility was greatly reduced.8

CORNEAL SENSITIVITY

Corneal sensitivity was measured by using a gas esthesiometer11 operated by an observer who was masked to the treatment received by each eye. A mechanical stimulus was provided by a stream of air directed at the center of the cornea through a nozzle with an inside diameter of 0.5 mm, positioned 5 mm from the central cornea. The stimulus flow rate ranged from 25 to 270 mL/min, and the air was warmed to minimize the effects of evaporative cooling on the corneal surface. Stimuli were administered to the central cornea by turning on the airflow for 2 seconds. After each stimulus, the participant reported whether he or she experienced a sensation. A superthreshold stimulus (150 mL/min) was administered first. Flow rate was then reduced to 25 mL/min, a subthreshold rate, and stimuli were then administered at flow rates that increased in steps of 25 mL/min until the participant reported a sensation. The next stimulus was administered at 12 mL/min less than the detected stimulus. If the participant felt this stimulus, we assumed that the threshold was at a flow rate 6 mL/min below this rate; if the patient did not feel it, we assumed that the threshold was 6 mL/min above this rate. A null stimulus (no airflow) was administered at least once during each trial to ensure that participants did not incorrectly respond (false-positive). Measurement of left and right eyes was alternated between stimulus presentations, maintaining a minimum 2-minute interval between stimulus administrations in each eye. If the participant did not feel the initial stimulus at 150 mL/min, the next stimulus was increased to 250 mL/min. If this was felt, the next stimulus was administered at 175 mL/min, and subsequent stimuli were increased in 25 mL/min steps as described.

During the 3-year study, we noticed a significant change in mechanical threshold of control eyes at 36 months compared with previous examinations. As a result, in addition to reporting the absolute mechanical thresholds and comparing absolute mechanical thresholds in LASIK eyes with those in control eyes, we calculated a mechanical threshold ratio, which was the ratio of the mechanical threshold in LASIK eyes to the mean mechanical threshold of control eyes at the concurrent examination. Unlike the LASIK participants, controls were not examined at 3 and 6 months. Thus, the mean mechanical threshold of control eyes at 1 month was used for comparison with the absolute thresholds at 3 and 6 months to calculate the threshold ratio at 3 and 6 months. A mechanical threshold ratio higher than 1 indicated lower corneal sensitivity of LASIK eyes relative to controls, and a mechanical threshold ratio less than 1 indicated higher corneal sensitivity of LASIK eyes relative to controls.

STATISTICAL ANALYSIS

Differences between eyes (ie, treatments) at each examination and differences between preoperative and postoperative examinations for each treatment were assessed by using 2-tailed paired t tests if the data were distributed normally and Wilcoxon signed rank tests if they were not. Differences between treatments and differences between the postoperative examinations compared with preoperative examinations within each treatment were adjusted for multiple comparisons by using the Bonferroni method. Differences in mechanical threshold after LASIK were also assessed for all eyes, regardless of the method of flap creation, by using generalized estimating equation models to account for possible correlation between fellow eyes of the same participant.12 Differences between LASIK eyes and controls eyes were also assessed by using generalized estimating equation models. P ≤ .05 was considered statistically significant.

The trial was powered to assess differences in visual acuity13 and corneal haze7 between treatments. Corneal nerve density and sensitivity were secondary outcomes, and thus post hoc power analyses were performed where appropriate by calculating minimum detectable differences for nonsignificant differences. For minimum detectable differences, we assumed that there were 21 independent observations (α = .05/5 or .05/6 depending on the comparison, β = .20). All data are reported as mean (SD) unless otherwise indicated.

PARTICIPANTS

Confocal microscopy data were unavailable for both eyes of 1 participant at 36 months after LASIK. Four eyes of 2 participants required LASIK enhancement procedures for mild undercorrections, which were similar in the fellow eyes; the enhancements occurred soon after the 12-month examination, and data for these eyes were retained for analysis at 36 months. One eye of 1 patient experienced trauma-induced recurrent erosions between 13 and 22 months after surgery; no erosions occurred after that time, and sensitivity data for this eye were included at 36 months but confocal microscopy was not performed.

In the control group, 1 participant did not return for examination at 12 months. Also, 2 participants from that group did not return at 36 months.

FLAP VARIABLES

Measured central flap thickness at 1 month was 143 (16) μm with the femtosecond laser and 138 (22) μm with the microkeratome.7 The actual diameter of bladeless flaps, as reported by the femtosecond laser, was 8.9 (0.2) mm, and the intended diameter of microkeratome flaps was 9.3 (0.4) mm (the actual diameter of the microkeratome flaps was not measured); the mean difference was 0.4 (0.4) mm (P < .001).

SUBBASAL NERVE DENSITY

Subbasal nerve density did not differ between femtosecond and microkeratome treatments at any examination before or after LASIK (Figure 1 and Table 1). For both treatments, subbasal nerve density was decreased at 1 month after LASIK (P < .001) and remained decreased through 12 months (P < .001) (Table 1). At 36 months, subbasal nerve density did not differ from that of the preoperative examination for either treatment, but the minimum detectable differences were 7081 and 7930 μm/mm2 for the femtosecond laser and mechanical microkeratome, respectively (α = .05/5, β = .20, paired analyses).

Place holder to copy figure label and caption
Figure 1.

Corneal subbasal nerve density before and after laser in situ keratomileusis (LASIK). Subbasal nerve density did not differ between femtosecond laser (bladeless) and mechanical microkeratome treatments at any examination before or after LASIK. For both treatments, subbasal nerve density was decreased at 1 month after LASIK (P < .001) and remained decreased through 12 months (P < .001). At 36 months, subbasal nerve density did not differ from the preoperative density for either treatment; minimum detectable differences were 7081 μm/mm2 and 7930 μm/mm2 for femtosecond laser and mechanical microkeratome, respectively (α = .05/5, β = .20, paired analyses).

Graphic Jump Location
Table Graphic Jump LocationTable 1. Subbasal Nerve Density Before and After LASIKa
CORNEAL SENSITIVITY

Absolute mechanical thresholds did not differ between femtosecond and microkeratome treatments at any examination before or after LASIK and did not differ at any examination after LASIK compared with before LASIK within each treatment (Figure 2 and Table 2). In the concurrent controls, absolute mechanical threshold remained stable except for an increase at 36 months compared with 12 months (P = .01; Figure 2 and Table 2). As a result, we calculated mechanical threshold ratios for the LASIK eyes and found no difference between treatments at any examination and no difference between preoperative and postoperative values within treatments (Figure 3 and Table 2). Absolute mechanical thresholds also did not differ between LASIK and control eyes at any examination (P ≥ .21).

Place holder to copy figure label and caption
Figure 2.

Absolute mechanical threshold before and after laser in situ keratomileusis (LASIK) and in concurrent control corneas. Central corneal mechanical thresholds (inversely related to corneal sensitivity) did not differ between femtosecond laser (bladeless) and mechanical microkeratome treatments before or after LASIK. The mechanical threshold appeared to increase transiently at 1 month after LASIK, but the threshold was not significantly greater than it was before LASIK within each treatment. The increase in absolute mechanical threshold from 12 to 36 months after LASIK was accompanied by a similar increase in mechanical threshold in concurrent, unoperated-on control corneas.

Graphic Jump Location
Place holder to copy figure label and caption
Figure 3.

Mechanical threshold ratio before and after laser in situ keratomileusis (LASIK). The mechanical threshold ratio is the ratio of absolute mechanical threshold in corneas undergoing LASIK to absolute mechanical threshold in control corneas and was calculated to account for the drift in absolute mechanical threshold noted in control corneas between 12 and 36 months (see Figure 2). The mechanical threshold ratio did not differ between femtosecond laser (bladeless) and mechanical microkeratome treatments at any examination before or after LASIK. The transient increase in the mechanical threshold ratio at 1 month after LASIK was not statistically significant within each treatment group (femtosecond laser, P = .65; microkeratome, P = .39) but, when all eyes were combined, the increase from the preoperative ratio was significant (P = .05), ie, decreased sensitivity, and returned to the preoperative ratio by 3 months (P = .99) with no subsequent change.

Graphic Jump Location
Table Graphic Jump LocationTable 2. Absolute Threshold of Mechanical Sensitivity and Mechanical Threshold Ratioa

When fellow eyes of each participant were combined, the mechanical threshold ratio increased (decreased sensitivity) at 1 month (1.29 [0.85]) after LASIK compared with the preoperative value (0.89 [0.73], P = .05), returned to the preoperative value by 3 months (P = .99), and remained stable thereafter (42 eyes, generalized estimating equation analysis). When fellow LASIK eyes were combined, there were no differences in absolute mechanical thresholds compared with control eyes at any examination (Table 2) (P ≥ .24). Subbasal nerve density did not correlate with corneal sensitivity before or at any examination after LASIK.

Corneal subbasal nerve regeneration after LASIK did not differ between eyes with the flap created by the femtosecond laser compared with eyes with the flap created by the mechanical microkeratome. Similarly, the changes in corneal sensitivity after LASIK did not differ by treatment, and the recovery of corneal sensitivity did not correspond to subbasal nerve regeneration.

Corneal stromal nerves course anteriorly in the anterior stroma to pierce the Bowman layer and form the subbasal nerve plexus at the basal aspect of the basal epithelial cells.14 The nerves are devoid of myelin sheaths to aid corneal transparency and are surrounded by Schwann cells only.15 Stromal reinnervation after corneal wounding is facilitated by apposition of wound edges, and this might be explained by the realignment of proximal and distal Schwann cell channels.16,17 As a result, reinnervation after penetrating keratoplasty, in which Schwann cell channels do not align, is virtually absent.17,18 Central subbasal and stromal nerve fiber bundles are also virtually absent immediately after LASIK because of transection of nerves during flap creation and subsequent stromal photoablation.19 The nerves regrow after LASIK, although through 5 years their density remains less than it is preoperatively1; regrowth might be aided by good realignment of the Schwann cell channels at the flap margin. Femtosecond lasers create planar flaps with precise and defined side-cut angles; thus, the periphery of the flap is as thick as the center and has an almost vertical edge. This geometric configuration should allow more precise realignment of femtosecond laser flaps compared with microkeratome flaps, which have tapered edges. Nevertheless, our results showed that femtosecond laser flaps did not result in faster reinnervation. Our results were not confounded by creating flaps of different thicknesses because the measured flap thickness was similar between treatments (approximately 140 μm).7

Subbasal nerves were not visible in most eyes at 1 month after LASIK, and recovery of nerve density was slow, similar to that in previous studies.2023 Although we did not show a difference in subbasal nerve density at 36 months after LASIK compared with the preoperative state, the large standard deviation of nerve density and the small sample size prevented us from detecting differences smaller than approximately 7000 μm/mm2; previously, we found that subbasal density remained lower through 5 years after LASIK.1 Of note, subbasal nerve density in our previous study1 appeared lower than that in the present study because we used confocal microscopes with differences in their optical design.8,24

Corneal mechanical sensitivity did not differ between methods of flap creation at any examination. Lim et al25 suggested that corneal sensitivity recovered faster after LASIK with femtosecond laser flap creation compared with microkeratome flap creation; they found a difference in central sensitivity of 3-mm filament length by Cochet-Bonnet esthesiometry at 3 months in a small nonrandomized series of eyes. In our study, corneal mechanical sensitivity was decreased at 1 month after LASIK before returning to preoperative sensitivity by 3 months. Stapleton et al26 found similar results in a cross-sectional study with the same gas esthesiometer, and Darwish et al27 found similar results by using a different noncontact esthesiometer. Gallar et al28 used the same esthesiometer as was used in our study and concluded that there was a transient hyperesthesia within days after LASIK followed by reduced mechanical and chemical sensitivity for as long as 2 years after LASIK, although their data were not statistically different from those for controls. Although our absolute mechanical threshold data suggested a trend toward hyperesthesia at 3 to 12 months after LASIK compared with controls (Figure 2), the differences were not statistically significant.

Most other studies have used Cochet-Bonnet esthesiometry to measure corneal sensitivity after LASIK.29 With this method, Chuck et al30 found that corneal sensation transiently decreased after LASIK but returned to preoperative sensitivity by 3 weeks after surgery. In contrast, most studies reported decreased sensation through at least 6 months after LASIK,21,3134 and Mian et al35 showed decreased sensation throughout the first year after LASIK. Although Cochet-Bonnet esthesiometry is a crude method of measuring corneal sensitivity and is limited by ceiling and floor effects when corneal sensitivity is normal or very abnormal, it is a quick and easy test in clinical studies and has detected a reduction in corneal sensitivity after LASIK in several studies. Gas esthesiometry has been suggested to be more sensitive than Cochet-Bonnet esthesiometry for measuring mechanical sensitivity,11,28 but our results and those of Stapleton et al26 and Darwish et al27 suggest that gas esthesiometry might in fact be less sensitive than Cochet-Bonnet esthesiometry. The discrepancy might be explained by differences in the types of receptor and the geometric area stimulated by the 2 esthesiometry techniques.11,28,29 We also found that mechanical sensitivity significantly decreased between 12 and 36 months in the concurrent, unoperated-on control corneas; although this might be explained by interobserver variation because our observer changed between these examinations, the new observer used the same protocol as the previous one, and thus we cannot rule out a change in the gas esthesiometer. Had we not examined concurrent controls, we might have incorrectly concluded that corneal sensitivity decreased between 12 and 36 months after LASIK (Figure 2). With the gas esthesiometer, we found that the best control for an eye was its fellow eye (Table 2), and thus our conclusion of no difference in corneal sensitivity between methods of flap creation is valid.

The prolonged decrease in subbasal nerve density after LASIK was accompanied by only a transient decrease in corneal sensitivity at 1 month, and, in contrast to Lee et al,34 we did not find a correlation between subbasal nerve density and sensitivity at any examination. Although it is possible that the cornea is reinnervated by small-caliber nerves that are not visible with our confocal microscope,8 thus restoring sensation to the cornea, it is also possible that our inability to detect small changes in sensitivity after LASIK resulted in our inability to detect a relationship between nerve density and sensitivity.

Between 12 and 36 months, fellow eyes of 2 participants received LASIK enhancements, and we retained the data from these eyes at 36 months because the fellow eyes were treated for similar undercorrections, even though enhancements result in severing and ablation of corneal nerves and likely a change in corneal sensitivity. One eye of another participant experienced corneal erosions between the 12- and 36-month examinations; confocal microscopy was not performed, but the corneal sensitivity data were retained in the analysis at 36 months. We analyzed our data at 36 months after excluding the enhanced eyes and the eye with erosions (data not shown), and the results did not alter any of our conclusions.

In summary, we found that the method of flap creation did not affect the rate of corneal reinnervation or recovery of corneal sensitivity after LASIK. We were unable to find a relationship between corneal subbasal nerve density and corneal sensitivity.

Correspondence: Sanjay V. Patel, MD, Department of Ophthalmology, Mayo Clinic, 200 First St SW, Rochester, MN 55905 (patel.sanjay@mayo.edu).

Submitted for Publication: February 5, 2010; final revision received March 4, 2010; accepted March 10, 2010.

Financial Disclosure: Dr Bourne reports serving as a consultant for Allergan Inc.

Funding/Support: This study was supported in part by grant EY 02037 from the National Institutes of Health, Research to Prevent Blindness (with Dr Patel as an Olga Keith Wiess Special Scholar and an unrestricted grant to the Department of Ophthalmology at Mayo Clinic), and the Mayo Foundation.

Previous Presentation: This study was presented in part as a scientific poster at the European Society of Cataract and Refractive Surgeons; September 12-16, 2009; Barcelona, Spain.

Erie  JC McLaren  JWHodge  DOBourne  WM Recovery of corneal subbasal nerve density after PRK and LASIK. Am J Ophthalmol 2005;140 (6) 1059- 1064
PubMed
Wilson  SE Laser in situ keratomileusis-induced (presumed) neurotrophic epitheliopathy. Ophthalmology 2001;108 (6) 1082- 1087
PubMed
Sugar  A Ultrafast (femtosecond) laser refractive surgery. Curr Opin Ophthalmol 2002;13 (4) 246- 249
PubMed
Salomão  MQAmbrósio  R  JrWilson  SE Dry eye associated with laser in situ keratomileusis: mechanical microkeratome versus femtosecond laser. J Cataract Refract Surg 2009;35 (10) 1756- 1760
PubMed
Medeiros  FWStapleton  WMHammel  JKrueger  RRNetto  MVWilson  SE Wavefront analysis comparison of LASIK outcomes with the femtosecond laser and mechanical microkeratomes. J Refract Surg 2007;23 (9) 880- 887
PubMed
von Jagow  BKohnen  T Corneal architecture of femtosecond laser and microkeratome flaps imaged by anterior segment optical coherence tomography. J Cataract Refract Surg 2009;35 (1) 35- 41
PubMed
Patel  SVMaguire  LJ McLaren  JWHodge  DOBourne  WM Femtosecond laser versus mechanical microkeratome for LASIK: a randomized controlled study. Ophthalmology 2007;114 (8) 1482- 1490
PubMed
Erie  EA McLaren  JWKittleson  KMPatel  SVErie  JCBourne  WM Corneal subbasal nerve density: a comparison of two confocal microscopes. Eye Contact Lens 2008;34 (6) 322- 325
PubMed
Meijering  EJacob  MSarria  JCFSteiner  PHirling  HUnser  M Design and validation of a tool for neurite tracing and analysis in fluorescence microscopy images. Cytometry A 2004;58 (2) 167- 176
PubMed
Rasband  WS ImageJ, 1997-2009. US National Institutes of Health Web site. http://rsb.info.nih.gov/ij/. Accessed February 1, 2009
Belmonte  CAcosta  MCSchmelz  MGallar  J Measurement of corneal sensitivity to mechanical and chemical stimulation with a CO2 esthesiometer. Invest Ophthalmol Vis Sci 1999;40 (2) 513- 519
PubMed
Zeger  SLLiang  KY Longitudinal data analysis for discrete and continuous outcomes. Biometrics 1986;42 (1) 121- 130
PubMed
Calvo  R McLaren  JWHodge  DOBourne  WMPatel  SV Corneal aberrations and visual acuity after laser in situ keratomileusis: femtosecond laser versus mechanical microkeratome. Am J Ophthalmol 2010;149 (5) 785- 793
PubMed
Müller  LJVrensen  GFPels  LCardozo  BNWillekens  B Architecture of human corneal nerves. Invest Ophthalmol Vis Sci 1997;38 (5) 985- 994
PubMed
Müller  LJMarfurt  CFKruse  FTervo  TM Corneal nerves: structure, contents and function. Exp Eye Res 2003;76 (5) 521- 542
PubMed
Chan-Ling  TTervo  KTervo  TVannas  AHolden  BAEranko  L Long-term neural regeneration in the rabbit following 180° limbal incision. Invest Ophthalmol Vis Sci 1987;28 (12) 2083- 2088
PubMed
Tervo  TVannas  ATervo  KHolden  BA Histochemical evidence of limited reinnervation of human corneal grafts. Acta Ophthalmol (Copenh) 1985;63 (2) 207- 214
PubMed
Patel  SVErie  JC McLaren  JWBourne  WM Keratocyte density and recovery of subbasal nerves after penetrating keratoplasty and in late endothelial failure. Arch Ophthalmol 2007;125 (12) 1693- 1698
PubMed
Calvillo  MP McLaren  JWHodge  DOBourne  WM Corneal reinnervation after LASIK: prospective 3-year longitudinal study. Invest Ophthalmol Vis Sci 2004;45 (11) 3991- 3996
PubMed
Lee  BH McLaren  JWErie  JCHodge  DOBourne  WM Reinnervation in the cornea after LASIK. Invest Ophthalmol Vis Sci 2002;43 (12) 3660- 3664
PubMed
Bragheeth  MADua  HS Corneal sensation after myopic and hyperopic LASIK: clinical and confocal microscopic study. Br J Ophthalmol 2005;89 (5) 580- 585
PubMed
Linna  TUVesaluoma  MHPérez-Santonja  JJPetroll  WMAlió  JLTervo  TM Effect of myopic LASIK on corneal sensitivity and morphology of subbasal nerves. Invest Ophthalmol Vis Sci 2000;41 (2) 393- 397
PubMed
Moilanen  JAOHolopainen  JMVesaluoma  MHTervo  TMT Corneal recovery after LASIK for high myopia: a 2-year prospective confocal microscopic study. Br J Ophthalmol 2008;92 (10) 1397- 1402
PubMed
Erie  JC McLaren  JWPatel  SV Confocal microscopy in ophthalmology. Am J Ophthalmol 2009;148 (5) 639- 646
PubMed
Lim  TYang  SKim  MTchah  H Comparison of the IntraLase femtosecond laser and mechanical microkeratome for laser in situ keratomileusis. Am J Ophthalmol 2006;141 (5) 833- 839
PubMed
Stapleton  FHayward  KBBachand  N  et al.  Evaluation of corneal sensitivity to mechanical and chemical stimuli after LASIK: a pilot study. Eye Contact Lens 2006;32 (2) 88- 93
PubMed
Darwish  TBrahma  AO’Donnell  CEfron  N Subbasal nerve fiber regeneration after LASIK and LASEK assessed by noncontact esthesiometry and in vivo confocal microscopy: prospective study. J Cataract Refract Surg 2007;33 (9) 1515- 1521
PubMed
Gallar  JAcosta  MCMoilanen  JAHolopainen  JMBelmonte  CTervo  TM Recovery of corneal sensitivity to mechanical and chemical stimulation after laser in situ keratomileusis. J Refract Surg 2004;20 (3) 229- 235
PubMed
Cochet  PBonnet  R L’esthesie corneene: sa mesure clinique: ses variations physiologiques et pathologiques. Clin Ophthalmol 1960;43- 27
Chuck  RSQuiros  PAPerez  AC McDonnell  PJ Corneal sensation after laser in situ keratomileusis. J Cataract Refract Surg 2000;26 (3) 337- 339
PubMed
Kim  W-SKim  J-S Change in corneal sensitivity following laser in situ keratomileusis. J Cataract Refract Surg 1999;25 (3) 368- 373
PubMed
Donnenfeld  EDEhrenhaus  MSolomon  RMazurek  JRozell  JCPerry  HD Effect of hinge width on corneal sensation and dry eye after laser in situ keratomileusis. J Cataract Refract Surg 2004;30 (4) 790- 797
PubMed
Donnenfeld  EDSolomon  KPerry  HD  et al.  The effect of hinge position on corneal sensation and dry eye after LASIK. Ophthalmology 2003;110 (5) 1023- 1030
PubMed
Lee  SJKim  JKSeo  KYKim  EKLee  HK Comparison of corneal nerve regeneration and sensitivity between LASIK and laser epithelial keratomileusis (LASEK). Am J Ophthalmol 2006;141 (6) 1009- 1015
PubMed
Mian  SILi  AYDutta  SMusch  DCShtein  RM Dry eyes and corneal sensation after laser in situ keratomileusis with femtosecond laser flap creation: effect of hinge position, hinge angle, and flap thickness. J Cataract Refract Surg 2009;35 (12) 2092- 2098
PubMed

Figures

Place holder to copy figure label and caption
Figure 1.

Corneal subbasal nerve density before and after laser in situ keratomileusis (LASIK). Subbasal nerve density did not differ between femtosecond laser (bladeless) and mechanical microkeratome treatments at any examination before or after LASIK. For both treatments, subbasal nerve density was decreased at 1 month after LASIK (P < .001) and remained decreased through 12 months (P < .001). At 36 months, subbasal nerve density did not differ from the preoperative density for either treatment; minimum detectable differences were 7081 μm/mm2 and 7930 μm/mm2 for femtosecond laser and mechanical microkeratome, respectively (α = .05/5, β = .20, paired analyses).

Graphic Jump Location
Place holder to copy figure label and caption
Figure 2.

Absolute mechanical threshold before and after laser in situ keratomileusis (LASIK) and in concurrent control corneas. Central corneal mechanical thresholds (inversely related to corneal sensitivity) did not differ between femtosecond laser (bladeless) and mechanical microkeratome treatments before or after LASIK. The mechanical threshold appeared to increase transiently at 1 month after LASIK, but the threshold was not significantly greater than it was before LASIK within each treatment. The increase in absolute mechanical threshold from 12 to 36 months after LASIK was accompanied by a similar increase in mechanical threshold in concurrent, unoperated-on control corneas.

Graphic Jump Location
Place holder to copy figure label and caption
Figure 3.

Mechanical threshold ratio before and after laser in situ keratomileusis (LASIK). The mechanical threshold ratio is the ratio of absolute mechanical threshold in corneas undergoing LASIK to absolute mechanical threshold in control corneas and was calculated to account for the drift in absolute mechanical threshold noted in control corneas between 12 and 36 months (see Figure 2). The mechanical threshold ratio did not differ between femtosecond laser (bladeless) and mechanical microkeratome treatments at any examination before or after LASIK. The transient increase in the mechanical threshold ratio at 1 month after LASIK was not statistically significant within each treatment group (femtosecond laser, P = .65; microkeratome, P = .39) but, when all eyes were combined, the increase from the preoperative ratio was significant (P = .05), ie, decreased sensitivity, and returned to the preoperative ratio by 3 months (P = .99) with no subsequent change.

Graphic Jump Location

Tables

Table Graphic Jump LocationTable 1. Subbasal Nerve Density Before and After LASIKa
Table Graphic Jump LocationTable 2. Absolute Threshold of Mechanical Sensitivity and Mechanical Threshold Ratioa

References

Erie  JC McLaren  JWHodge  DOBourne  WM Recovery of corneal subbasal nerve density after PRK and LASIK. Am J Ophthalmol 2005;140 (6) 1059- 1064
PubMed
Wilson  SE Laser in situ keratomileusis-induced (presumed) neurotrophic epitheliopathy. Ophthalmology 2001;108 (6) 1082- 1087
PubMed
Sugar  A Ultrafast (femtosecond) laser refractive surgery. Curr Opin Ophthalmol 2002;13 (4) 246- 249
PubMed
Salomão  MQAmbrósio  R  JrWilson  SE Dry eye associated with laser in situ keratomileusis: mechanical microkeratome versus femtosecond laser. J Cataract Refract Surg 2009;35 (10) 1756- 1760
PubMed
Medeiros  FWStapleton  WMHammel  JKrueger  RRNetto  MVWilson  SE Wavefront analysis comparison of LASIK outcomes with the femtosecond laser and mechanical microkeratomes. J Refract Surg 2007;23 (9) 880- 887
PubMed
von Jagow  BKohnen  T Corneal architecture of femtosecond laser and microkeratome flaps imaged by anterior segment optical coherence tomography. J Cataract Refract Surg 2009;35 (1) 35- 41
PubMed
Patel  SVMaguire  LJ McLaren  JWHodge  DOBourne  WM Femtosecond laser versus mechanical microkeratome for LASIK: a randomized controlled study. Ophthalmology 2007;114 (8) 1482- 1490
PubMed
Erie  EA McLaren  JWKittleson  KMPatel  SVErie  JCBourne  WM Corneal subbasal nerve density: a comparison of two confocal microscopes. Eye Contact Lens 2008;34 (6) 322- 325
PubMed
Meijering  EJacob  MSarria  JCFSteiner  PHirling  HUnser  M Design and validation of a tool for neurite tracing and analysis in fluorescence microscopy images. Cytometry A 2004;58 (2) 167- 176
PubMed
Rasband  WS ImageJ, 1997-2009. US National Institutes of Health Web site. http://rsb.info.nih.gov/ij/. Accessed February 1, 2009
Belmonte  CAcosta  MCSchmelz  MGallar  J Measurement of corneal sensitivity to mechanical and chemical stimulation with a CO2 esthesiometer. Invest Ophthalmol Vis Sci 1999;40 (2) 513- 519
PubMed
Zeger  SLLiang  KY Longitudinal data analysis for discrete and continuous outcomes. Biometrics 1986;42 (1) 121- 130
PubMed
Calvo  R McLaren  JWHodge  DOBourne  WMPatel  SV Corneal aberrations and visual acuity after laser in situ keratomileusis: femtosecond laser versus mechanical microkeratome. Am J Ophthalmol 2010;149 (5) 785- 793
PubMed
Müller  LJVrensen  GFPels  LCardozo  BNWillekens  B Architecture of human corneal nerves. Invest Ophthalmol Vis Sci 1997;38 (5) 985- 994
PubMed
Müller  LJMarfurt  CFKruse  FTervo  TM Corneal nerves: structure, contents and function. Exp Eye Res 2003;76 (5) 521- 542
PubMed
Chan-Ling  TTervo  KTervo  TVannas  AHolden  BAEranko  L Long-term neural regeneration in the rabbit following 180° limbal incision. Invest Ophthalmol Vis Sci 1987;28 (12) 2083- 2088
PubMed
Tervo  TVannas  ATervo  KHolden  BA Histochemical evidence of limited reinnervation of human corneal grafts. Acta Ophthalmol (Copenh) 1985;63 (2) 207- 214
PubMed
Patel  SVErie  JC McLaren  JWBourne  WM Keratocyte density and recovery of subbasal nerves after penetrating keratoplasty and in late endothelial failure. Arch Ophthalmol 2007;125 (12) 1693- 1698
PubMed
Calvillo  MP McLaren  JWHodge  DOBourne  WM Corneal reinnervation after LASIK: prospective 3-year longitudinal study. Invest Ophthalmol Vis Sci 2004;45 (11) 3991- 3996
PubMed
Lee  BH McLaren  JWErie  JCHodge  DOBourne  WM Reinnervation in the cornea after LASIK. Invest Ophthalmol Vis Sci 2002;43 (12) 3660- 3664
PubMed
Bragheeth  MADua  HS Corneal sensation after myopic and hyperopic LASIK: clinical and confocal microscopic study. Br J Ophthalmol 2005;89 (5) 580- 585
PubMed
Linna  TUVesaluoma  MHPérez-Santonja  JJPetroll  WMAlió  JLTervo  TM Effect of myopic LASIK on corneal sensitivity and morphology of subbasal nerves. Invest Ophthalmol Vis Sci 2000;41 (2) 393- 397
PubMed
Moilanen  JAOHolopainen  JMVesaluoma  MHTervo  TMT Corneal recovery after LASIK for high myopia: a 2-year prospective confocal microscopic study. Br J Ophthalmol 2008;92 (10) 1397- 1402
PubMed
Erie  JC McLaren  JWPatel  SV Confocal microscopy in ophthalmology. Am J Ophthalmol 2009;148 (5) 639- 646
PubMed
Lim  TYang  SKim  MTchah  H Comparison of the IntraLase femtosecond laser and mechanical microkeratome for laser in situ keratomileusis. Am J Ophthalmol 2006;141 (5) 833- 839
PubMed
Stapleton  FHayward  KBBachand  N  et al.  Evaluation of corneal sensitivity to mechanical and chemical stimuli after LASIK: a pilot study. Eye Contact Lens 2006;32 (2) 88- 93
PubMed
Darwish  TBrahma  AO’Donnell  CEfron  N Subbasal nerve fiber regeneration after LASIK and LASEK assessed by noncontact esthesiometry and in vivo confocal microscopy: prospective study. J Cataract Refract Surg 2007;33 (9) 1515- 1521
PubMed
Gallar  JAcosta  MCMoilanen  JAHolopainen  JMBelmonte  CTervo  TM Recovery of corneal sensitivity to mechanical and chemical stimulation after laser in situ keratomileusis. J Refract Surg 2004;20 (3) 229- 235
PubMed
Cochet  PBonnet  R L’esthesie corneene: sa mesure clinique: ses variations physiologiques et pathologiques. Clin Ophthalmol 1960;43- 27
Chuck  RSQuiros  PAPerez  AC McDonnell  PJ Corneal sensation after laser in situ keratomileusis. J Cataract Refract Surg 2000;26 (3) 337- 339
PubMed
Kim  W-SKim  J-S Change in corneal sensitivity following laser in situ keratomileusis. J Cataract Refract Surg 1999;25 (3) 368- 373
PubMed
Donnenfeld  EDEhrenhaus  MSolomon  RMazurek  JRozell  JCPerry  HD Effect of hinge width on corneal sensation and dry eye after laser in situ keratomileusis. J Cataract Refract Surg 2004;30 (4) 790- 797
PubMed
Donnenfeld  EDSolomon  KPerry  HD  et al.  The effect of hinge position on corneal sensation and dry eye after LASIK. Ophthalmology 2003;110 (5) 1023- 1030
PubMed
Lee  SJKim  JKSeo  KYKim  EKLee  HK Comparison of corneal nerve regeneration and sensitivity between LASIK and laser epithelial keratomileusis (LASEK). Am J Ophthalmol 2006;141 (6) 1009- 1015
PubMed
Mian  SILi  AYDutta  SMusch  DCShtein  RM Dry eyes and corneal sensation after laser in situ keratomileusis with femtosecond laser flap creation: effect of hinge position, hinge angle, and flap thickness. J Cataract Refract Surg 2009;35 (12) 2092- 2098
PubMed

Correspondence

CME
Meets CME requirements for:
Browse CME for all U.S. States
Accreditation Information
The American Medical Association is accredited by the Accreditation Council for Continuing Medical Education to provide continuing medical education for physicians. The AMA designates this journal-based CME activity for a maximum of 1 AMA PRA Category 1 CreditTM per course. Physicians should claim only the credit commensurate with the extent of their participation in the activity. Physicians who complete the CME course and score at least 80% correct on the quiz are eligible for AMA PRA Category 1 CreditTM.
Note: You must get at least of the answers correct to pass this quiz.
You have not filled in all the answers to complete this quiz
The following questions were not answered:
Sorry, you have unsuccessfully completed this CME quiz with a score of
The following questions were not answered correctly:
Commitment to Change (optional):
Indicate what change(s) you will implement in your practice, if any, based on this CME course.
Your quiz results:
The filled radio buttons indicate your responses. The preferred responses are highlighted
For CME Course: A Proposed Model for Initial Assessment and Management of Acute Heart Failure Syndromes
Indicate what changes(s) you will implement in your practice, if any, based on this CME course.
NOTE:
Citing articles are presented as examples only. In non-demo SCM6 implementation, integration with CrossRef’s "Cited By" API will populate this tab (http://www.crossref.org/citedby.html).
Submit a Comment

Multimedia

Some tools below are only available to our subscribers or users with an online account.

Related Content

Customize your page view by dragging & repositioning the boxes below.

Articles Related By Topic
Related Topics
PubMed Articles